Immersion Type Water Bath?s Temp. Controller.
Various Functions from "Fixed Temp. Operation" to "Programmable Operation".
Data Output, External Communication, etc. are available as Option.
Specifications
Type
Basic Type
Multi Function Type
Oil/Water Compatible Type
Model
BF200
BF400
BF500
BF600
Performance
Temp. Setting Range
-20.0 to 100deg.C
0 to 200deg.C
Operating Temp. Range
Room Temp. +5 to 80deg.C
Room Temp. +5 to 180deg.C*1
Temp. Control Accuracy
0.05 to 0.1deg.C
0.02 to 0.05deg.C
0.05 to 0.2deg.C*1(*1 Condition: Setting Temp. 100deg.C, Testing Bath BZ100D, Silicon Oil Stickiness 50cst)
Components
Stirrer
Propeller Stirring
Water Jet Stirring
Propeller Stirring
Heater
Stainless Pipe Heater
1.0kW
1.2kW
Temp. Control System
PID Control by Micro Computer
Temp. Setting
Digital Setting by Key
Temp. Display
Digital Display by Green LED
Min. Digit Indication: 0.1deg.C
Min. Digit Indication: 1deg.C
Setting Temp./ Measured Temp. Changeable
Indicated on Main Indicator(Sub-indicator: Indicate Setting Temp.)
Timer
Nil
1min. to 99h.59min. or 100h. to 999h.
Timer Resolution
Nil
1min. or 1h.
Operating Function
Fixed Operation
Fixed Operation: Continuous Operation by Set Temp., Quick Automatic StopProgramming Operation: 1 to 3 Optional Pattern, Max. 5 Segment Pattern (Repeating, Gradient Operation)
Additional Function
Temp. Pre-setting(Memory/Recall 1Temp.)
Temp. Pre-setting (Memory/Recall 10 Temp.)Timer Function (to 49,999h.), Key Lock Function, Water Jet Strength Changeable Function (10 steps Changeable)
N/A
Pump Jet Angle Changeable
N/A
N/A
N/A
Temp. Output (10mV/deg.C),External Alarm Output,Time ?up Output, External Communication Function (RS422A)
N/A
Heater Circuit Control
Triac Zero-cross Type
Sensor
Thermister
Pt Sensor
Safety Device
Safety Countermeasure
Shelf Diagnosis Function (Automatic Overheat Prevention, Abnormal Temp. Sensor,Heater Wire Broken*2, Triac Short Circuit*2, Main Relay Defect*2), Circuit Protector, Float Type Empty Boiling Preventor, Buzzer Alarm for Abnormal Condition (*2: This Function is Not Applicable for BF200)
Accessories
Testing Bath Model BY100 (Polypropylene)
Standard
External Dimension
W140 x D138 x H305mm
Clamp Available Thickness
Max. 35mm
Power Source
AC 100V 50/60Hz 11A
AC100V 50/60Hz 13A
Weight
Approx. 7kg
Model
BF200
BF400
BF500
BF600
Digital Setting/Display
Temp. Pre-setting
Programming Function
-
Water Jet Strength Changeable Function
-
-
Propeller Stirring
-
-
External Output, Communication Function, etc.
-
-
-
Oil Temp. Control
-
-
-